Bi-Functional Coated Tapered LPFG Sensor: Gas and Temperature Sensing

Bi-Functional Coated Tapered LPFG Sensor: Gas and Temperature Sensing

Zahraa S. AlshaikhliWasan A. Hekmat Hui Wang 

Laser and Optoelectronics Engineering Department, University of Technology, Baghdad 10066, Iraq

School of Biomedical Engineering, University of Western Ontario, London, Ontario N6A 3K7, Canada

Corresponding Author Email: 
zahraa.s.qasim@uotechnology.edu.iq
Page: 
831-838
|
DOI: 
https://doi.org/10.18280/mmep.090333
Received: 
20 January 2022
|
Revised: 
13 April 2022
|
Accepted: 
22 April 2022
|
Available online: 
30 June 2022
| Citation

© 2022 IIETA. This article is published by IIETA and is licensed under the CC BY 4.0 license (http://creativecommons.org/licenses/by/4.0/).

OPEN ACCESS

Abstract: 

A bi-functional coated tapered Long Period Fiber Grating LPFG as a gas and a temperature sensor is proposed. Obtains a sensor with a high sensitivity is still a matter of challenge. In this work, a taper LPFG is used for sensing purposes. Besides, in order to increase the sensor performance, numbers of Cr and Al layers are coated on the taper region. In this work, different concentrations of four gases Ar, NH3, CH4 and N2 are applied for temperature range (25℃ up to 80℃). however, the experimental results illustrate that the transmission loss is reduce which means a significant interaction between sensing medium and cladding mode is occurred successfully which leads to higher sensitivity. Nevertheless, from the main findings, the temperature sensitivities are 2.9 pm/℃, 4.9 pm/℃, 4.5 pm/℃ and 3.1 pm/℃ for 4% concentrations of Ar, NH3, CH4 and N2 respectively. In addition, the gases sensitives are 0.213 nm%-1, 0.41 nm%-1, 0.39 nm%-1, 0.201 nm%-1 with limit of detection LoD of 0.061%, 0.024%, 0.032% and 0.073% for Ar, NH3, CH4 and N2 respectively. Additionally, the sensor shows a good repeatability, recoverability and stability. In terms of stability, the sensor shows a stable value of wavelength shift under different values of gas concentration and temperatures which was estimated 1.55 µm. Besides, it shows a fast responsivity with 53 s and 60 s as response time and recovery time respectively. Interestingly, with less than 0.06 nm of standard deviation, the signal returned to its original resonance wavelength which confirms that this sensor has a good recoverability. From these results, a coated taper LPFG can be a promising candidate for medical and industrial applications.

Keywords: 

taper optical fiber, metal coated optical fiber, LPFG, taper LPFG, gas sensor, temperature sensor

1. Introduction

Temperature and gas sensors have been used widely in many industrial applications. Fiber optic based grating sensor has been utilized in many functions instead of conventional sensors due to its low cost, low weight, immunity to electromagnetic field, small size and its high sensing ability to different physical parameters such as temperature [1], pressure [2], acceleration [3], force [4], strain [5], displacement [6], high magnetic field [7] and chemical application [8]. Therefore, Fiber Bragg Grating FBG (few of hundred micrometers grating periods) and Long Period Fiber Grating LPFG (few of hundred nanometers grating periods) have been used for sensing several measurands. As a consequence, many studies have been conducted using FBG and LPFG on different fiber structures and different combination of FBG and LPFG by coating it with different materials [9]. For example, one study used five different apodization profiles based on FBG temperature sensor such as uniform, raised sine, sinc [10]. The study concluded that Nuttall profile gives a higher sensitivity of 0.01372 nm/℃ at 117.3℃. Another study proposed a S-shape taper optical fiber based on Mach Zehnder interferometer; the study showed that at range of 28-32℃, the temperature sensitivity was -15.66 nm/℃ [11]. A very recent study presented a hybrid structure of fiber optic based on a thin taper and an air bubble for temperature, refractive index and transverse load sensitivities [12]. However, the air bubble represents a Fabry Perot and Mach Zehnder interferometers; this structure gives 1.596 nm/N, 103.2 pm/℃ and -103.2 nm/RIU for transverses load, temperature and refractive index sensitivities respectively. D-shaped fiber based on LPFG as a strain and temperature sensor proposed; the etching and re-modulation method has a direct impact on the refractive index distribution which gives different coupling coefficients [13]. This structural gives a 10.75 pm/µɛ and 0.14 pm/℃ for strain and temperature sensitivities respectively. Further explore on D-shape based on LPFG sensor was carried out for bending measurements [14]. This novel structure formed by splicing between multi-mode and single mode optical fiber. This sensor had the ability to discriminate the bending direction due to the asymmetric refraction index modulation. Moreover, a study illustrated that taper fiber based on FBG can be used in sensor of microsurgical force [15]. Another research approved that using taper FBG in medical field to identify cancers efficiently [16].

By summarizing recent researches, it is found that taper fiber structure used for sensing application more than other structure and it takes a significant attention by researchers because of its high ability for sensing through the sensing taper region. Further studies on enhancing the senor sensitivity have been proposed by coating the sensor with different materials such as metal and polymer. For instance, one research presented an s-shape LPFG sensor coated by metal for magnetic field sensitivity [17]. The sensor shows that the metal layer has an impact on increasing the coupling coefficient and hence increases the transmission loss with increasing magnetic field. However, the sensitivity was estimated to be -0.0084 dB/Gauss [18]. In terms of polymer, one study proposed a temperature sensor based on LPFG coated by Poly-dimethylsiloxane (PDMS). The study confirmed that the sensitivity enhanced by four times that bare LPFG. The sensitivity reached to 255.4 pm/℃ in the range of (20-80℃) [18]. Another interesting study combined FBG with TiO2 coated LPFG for trucking organic solvents [19, 20]. However, more methods of LPFG and FBG combination have been proposed. For instance, a study used LPFG followed by FBG as a refractometer based on a cladding mode Bragg grating (CBG) for a compensation of temperature. The authors claimed that this sensor is a promising candidate for biochemical detection [21]. Another study proposed a tapered optical fiber connected with photonic structure and it operates for 1.55 µm. Interestingly, this design has been used as a sensor and utilized it for maximizing the coupling efficiency to the photonic structure [22]. Similarly, a study showed that fabricated an LPFG with grating period of 400 µm and length of 5 cm in a biconical tapered fiber with 3.2 cm length and 34 µm diameter which was used effectively for refractive index sensing [23]. A novel research showed a high sensitivity performance of Pd-Ag coated layer on tapered sensor was used to measure hydrogen and ammonia. This novel sensor was able to sense a gas with less than 4% concentration [24]. A periodically taper optical fiber sensor (PTOF) was proposed as a temperature sensor for temperature values from 20℃ to 70℃, this study proved that the number of waists could have a significant impact on the temperature sensing [25]. Furthermore, another study used the same concept of periodically tapered sensor but for gas sensing. The authors reported that this sensor can be sufficiently utilized for in-situ fuel adulterations tracking [26]. A novel pressure sensor reported for the first time, this sensor is a periodic tapered structure that has a diameter ranging from 115 µm to 120 µm. the sensor gives an accurate pressure sensitivity equal to 5.1 pm/bar. This novel sensor fabricated by assembled equipment using arc discharge controlled by computer [27].

This study is proposed a dual function of temperature and gas sensitivity using tapered optical fiber based on LPFG coated by metal layer. Metal coating layer is selected for this study due to its advantages on increasing the temperature sensitivity.

2. Sensing Concept

As mentioned earlier, LPFG is periodic structure of several hundreds of micrometers grating. This periodic structure is a fiber core refractive index modulation which leads to light coupling between core and cladding modes. In general, the basic concept of optical fiber sensor based on grating is relying on wavelength shifting. In other words, wavelength shifting is induced by any changing in grating period and any modulation in the refractive index. However, sensing concept of LPFG is based on grating period changing Ʌ and modulation between refractive indices of nco and ncl. The following basic expression can summarize the concept in terms of Bragg wavelength λBLPG and grating period Ʌ [19]:

$\lambda_{c o}=2 n_{c o} \Lambda$      (1)

$\lambda_{c l}=2 n_{c l} \Lambda$       (2)

$\lambda_{B_{L P F G}}=\left(n_{c o}-n_{c l}\right) \Delta_{L P F G}$       (3)

where, (nco-ncl) ∆LPFG is the refractive index change of LPFG. The transmission spectrum modified if any external physical factors have applied on the sensor, especially, gas or temperature as it will explain in this work. This will lead to shifting in LPFG resonance wavelength and hence the transmission spectrum will change. In this work a thin layer of metal coating will be covered the tapered region of the sensor. This will have an impact on the sensitivity performance as it will be explaining the following sections. However, in terms of temperature sensitivity, the expression of transmission spectrum will be as follows taking into account the coating material [21].

$\Delta \lambda_{L P F G} / \Delta T_{L P F G}=\lambda_{L P F G}\left(\Delta_{n d} \alpha_{c o} \alpha_{\text {coating }}+2 \xi\right)+f(\Delta n)$        (4)

where, ∆λLPFG and TLPFGare wavelength change and temperature change for LPFG respectively. nd is (nco – ncl). f(∆n) is nonlinear function that explains the LPFG sensing for refractive index modulation. From Eq. (4), it is clear that the changing in LPFG resonance wavelength depending linearly on the changing in temperature and refractive index. According to gas concentration pressure sensitivity for LPFG can be expressed by the following equation:

$\Delta \lambda_{L P F G} / \Delta P_{L P F G}=\lambda_{L P F G}\left[-\frac{\left(1-2 \sigma_{c o}\right)}{E_{c o}}+\frac{n_{c o}^{2}}{2 E_{c o}}\right.$$\left.+\left(1-2 \sigma_{c o}\right)\left(2 p_{12}+p_{11}\right)-\frac{\sigma_{A l} r_{A l}^{2}}{E_{A l} \sigma_{A l}\left(r_{A l}^{2}-r_{c o}^{2}\right)}\right]$         (5)

where, σco, rco and Eco are core axial stress, radius and young’s modulus respectively. And p12 and p11 are the index of Pockel. Besides, EAL, σAL and rALare Young’s modulus, axial stress and radius Al layer respectively. This paper presents a bi-function tapered sensor for temperature and gas sensitivity coated by Al thin layer in order to improve the sensitivity performance.

3. Sensor Preparing Process

A silica Multi Mode Optical Fiber MMF was used in this study. It has 62 µm as a core diameter and 125 µm as a cladding diameter. The refractive indices are 1.45 and 1.44 for core and cladding respectively. The first step was preparing the taper region. A tapered region of optical fiber can be achieved by heating required position of the optical fiber and stretching it carefully. It can be heated by for example flame, arc discharge or laser radiation with high power. In addition, the taper region can be achieved by polishing and chemical etching [20]. As a result of these processes, the core of optical fiber become thinner along with the fiber itself. The thinner region can be few of millimeters. In the current work, the MMF was heated and stretched uniformly and its parameters were controlled based on the following expression which clearly explains the exponential shape transition of the optical fiber [20].

$d=d_{0} \exp (-z / 2 L)$        (6)

where, do is the overall MMF diameter and d is the diameter of waist region. Besides, z is the elongation distance and L is the waist length as shown in Figure 1.

Figure 1. Tapered optical fiber

The taper fabrication setup used in this work is similar to what was reported in Ref. [22]. A flame torch was fixed on a 3D transition stage controlled by motor, securing that the flame will apply uniformly on whole area of interest of optical fiber sensor. The two ends of optical fiber were fixed on linear transition stages which were fixed in opposite directions. One end of optical fiber was connected to Smart Scan Interrogator SSI and the second end connects to Optical Spectrum Analyzer OSA in order to monitor the changing in the transmission spectrum in real time as depicted in Figure 2.

Figure 2. Tapered MMF process

According to the grating inscription of Tapered Multi Mode Optical Fiber TMMF sensor, amplitude mask technique was used. This method involves on exposing the fiber core to excimer laser through amplitude mask, but firstly, increasing the photosensitivity of fiber core is essential step. For three days, our TMMF was exposed to Hydrogen aiming to increase the photosensitivity and hence, maximizing the rate of grating inscription. However, in this work, TMMF was exposed to 248 nm and power of 16 mJ of excimer laser KrF. A UV light was passed through an amplitude mask making a pattern of light and dark shades on the core. These patterns referred to the grating made on the fiber core. The grating period was 250 µm. Moreover, in order to increase the sensitivity performance, the TMMF will be coated by a metal of a thin layer of Al. A thermal vapor deposition was employed. Before Al coating, a thin layer of almost 3 nm of Cr was sputtered on the tapered region in order to maximize the adhesion rate of Al on tapered region. Besides, this step will prevent the Al layer from peeling off during the cleaning process. Then by using thermal vapor deposition, a layer of 19 nm of Al was deposited on tapered region. The Al layer thickness was monitored and controlled in real time by tracking the quartz crystal in deposition machine. It is important to highlight that the TMMF sensor was fixed on rotational platform to ensure that the metal deposited uniformly. The overall process of preparing the coated tapered LPFG sensor is shown in Figure 3.

Figure 3. Schematics of preparing process coated tapered LPFG sensor

The TMMF has the specifications of 15 mm as whole length, 8 mm as tapered region, 10 µm as waist diameter and 19 nm as coating thickness of Al. Besides, the operating wavelength was 1500 nm < λ < 1550 nm.

Figure 4. Schematics of bi-function sensor setup

4. Experimental Work

In order to test the sensor performance for gas and temperature sensitivity, a simple setup was employed for this purpose and it is shown in Figure 4. The sensor was fixed on two fixed stages with glued points. The sensing region was inside an aluminum tube wrapped with heating foil for temperature testing purpose. The heating foil was connected to heating system which consists of thermocouple, Pico Recorder PR, and power supply PS. The two ends of TMMF were linked to Smart Scan Interrogator SSI, OSA and PC. The sensing region setup was enclosed by cylindrical chamber with 1-liter volume for gas testing purpose. The applied temperatures were from 25℃ up to 80℃ while and for gas testing, four gases have been applied and controlled in this work Ar, NH3, CH4 and N2. The cylindrical chamber has two ports for gas inlet and gas outlet. However, by changing the applied voltage from PS, the temperature of heating foil arises and heats the surrounding environment of TMMF sensor. Besides, by adjusting the gas controller, the amount and type of applied gas are controlled.

5. Results and Discussion

5.1 Sensor characterization

Firstly, the sensor transmission spectrum before and after Al coating is shown in Figure 5. From the figure, it observes clearly that the spectrum extra emerged due to the impact of the metal layer (Al coating layer) which increased the sensitivity performance of the TMMF sensor. The spectrum has two loss peaks at approximately 1513 nm and 1536 nm which referred to successful inscription of LPFG on the core. Furthermore, the loss peak at 1536 nm has a transmission loss of -14 dB which is less than the loss of other loss peak hence has larger coupling efficiency compared to other peak. The minimum loss means that this loss peak produces a significant interaction between sensing medium and cladding mode hence produces significant sensitivity. In the following context, it has been concentrated only on the loss peak at 1536 nm in terms of the sensing performance. Moreover, the relation of different Al thicknesses with the wavelength shift has been shown in the inset Figure 5. At 19 nm thickness of Al, the sensor achieves the highest sensitivity. The final step was monitored the sensor by a Spectroscopic Ellipsometer in order to check the quality of deposition process.

Figure 5. Transmission spectrum of the TMMF before (black line) and after (red line) metal deposition. The impact of Al coated thickness on the wavelength shifting was shown in the inset

5.2 Sensor performance

Gas sensing performance was investigated for four gases Ar, NH3, CH4, and N2. For each gas four different gas concentrations were investigated. It is found that the TMMF sensor shows higher sensitivity for NH3 and CH4 than Ar and N2. The transmission spectrums are shown in Figure 6. The transmission spectrum confirms that after exposing the sensor to different gas concentrations, the transmission loss is incrementally reduced and the wavelength is shifted. For example, in 4% of NH3, the resonance wavelength shifts from 1536 nm to 1538.5 nm with significantly decreases in loss from -13 dB to -15 dB. Besides, these results confirm that the Al layer and the taper waist region contributed in improving the evanescent field intensity on the fiber core and hence, enhancing the fiber core mode conversion efficiency as many studies proven that [23]. In other words, when the gas concentration around the sensing region increase, the effective refractive index of Al layer is increased as well. This is because of the diffusion of the gas molecules into Al layer which caused improving the interaction between this region and evanescent field intensity and hence, transmission spectrum modified with less power loss. The sensor shows a similar performance scenario for CH4. In terms of Ar and N2 the sensor shows a small value of resonance wavelength shifts compared with other two gases which was estimated to be around 0.5 nm for 4% gases concentration.

A temperature TMMF sensor was experimentally investigated along with the impact of applying gas on it. The applied temperatures were from room temperature up to 80℃ at 4% of each Ar, NH3, CH4 and N2. Four temperature degrees were selected 30℃, 40℃, 50℃ and 75℃ in order to illustrate the transmission spectrum. Figure 7 shows how different temperatures can modify the transmission spectrum along with impact of gas concentrations. It can be seen that the sensor shows a significant responsivity to temperatures sensitivity. Figure 7 demonstrates the transmission spectrum under the effect of applied temperatures and gas concentrations which leads to a higher shift in the resonance wavelength. For example, for 4% NH3, the resonance wavelength shifts from 1536 nm to 1550.2 nm when the temperature reached 75℃. In addition to, the enhancement in the results belong to the existence of the metal layer which boosts the ability of the sensor to be more sensitive to small amount of temperature as many researches confirmed that [24]. Furthermore, the taper waist region contributes in positive aspect for enhancing the temperature sensitivity since, the applied temperatures will be direct to the sensing region which consist of the grating structure and the metal coating layer. This is why many researches have been preferred using tapered structure as a temperature sensor [20,21]. Nevertheless, the temperature sensitivities for 4% of Ar, NH3, CH4 and N2 were 2.9 pm/℃, 4.9 pm/℃, 4.5 pm/℃ and 3.1 pm/℃ respectively.

Moreover, to investigate the linearity performance of the TMMF sensor, the resonance wavelength shifts are estimated as a furcation of both gas concentrations and different temperatures which were extracted from Figures 6 and 7. These linear fit relations were plotted as Figure 8 illustrated. However, the sensor shows a good linearity performance under different gas concentrations. The correlation coefficients of 4% for each of Ar, NH3, CH4 and N2 were 0.995,0.998, 0.997 and 0.994 respectively; besides, the sensitivities for these gases are 0.213 nm%-1, 0.41 nm%-1, 0.39 nm%-1, 0.201 nm%-1 with congruent limit of detection LoD of 0.061%, 0.024%, 0.032% and 0.073% respectively. However, using metal layer with taper structures gives more enhanced linear sensor performance by factor 4 compared with [20, 24].

5.3 Sensor responsivity

The response characteristics of the TMMF sensor were tested further. For example, repeatability and stability tests were performed successfully. Experiments were carried out by applying 4% of NH3 on the TMMF sensor. The resonance wavelength shifts as a function of time for both gas and temperature sensors were plotted in Figure 9 and 10. In Figure 9a and for gas sensor, the stable value of wavelength shifts under 4% of NH3 was estimated to be 1.55 pm. The response time (tres) related to the TMMF sensor under the mentioned conditions was measured to be 53 s and the recovery time (trec) was calculated to be 60 s. Additionally, with less than 0.06 nm of standard deviation, the signal can return to its original resonance wavelength which confirms that this sensor has a good recoverability. According to the temperatures sensor, a repeatability test is shown in Figure 9b. The test was carried out at 4% of NH3 under temperatures range from room temperature up to 75℃. The time taken for two cycles was approximately 3 hours. Throughout this time, the sensor shows a good recoverability and repeatability under the same measurements conditions.

Figure 6. Transmission spectrum of the TMMF gas sensor (different gases concentrations)

Figure 7. Transmission spectrum of the TMMF temperature sensor

(a)

(b)

Figure 8. Resonance wavelength shift as a function of (a) gas concentrations at 25℃ and (b) temperatures changes

(a)

(b)

Figure 9. Repeatability response of TMMF as (a) gas sensor to 4% NH3 and (b) as a temperature sensor at 75℃

(a)

(b)

Figure 10. Stability test for TMMF sensor performance for (a) gas sensor at 25℃ and (b) temperature sensor at 4% of NH3

In terms of stability test of the TMMF sensor, Figure 10 illustrates the performance of the gas and temperature sensors. Firstly, the sensor was exposed to the three different concentrations of NH3 gas at room temperature for one hour, however, the sensor shows good stability measurements as shown in Figure 10a. The resonance wavelength shift was almost the same with an estimated percentage error of 1%, 2% and 2.5% for NH3 concentrations 1%, 3% and 4% respectively. Then, the sensor was exposed to fixed NH3 concentrations at different temperatures 30℃, 60℃ and 75℃ for one hour. Similarly, the sensor shows a good and similar wavelengths shift measurements with percentages error of 1% for all applied temperatures which indicates that the TMMF sensor has efficient performance as gas and temperature sensor at the same time; the test outcomes shown in Figure 10b. Again during these tests, it has been concentrated on keeping the performance of these tests under the same conditions. As a result, due to the metal coated layer (Cr and Al deposition) and the taper MMF structure, the sensor performance and sensor responsivity were enhanced significantly compared with other reported studies as mentioned earlier. Additionally, writing long period grating within the fiber tapered section could have impact on the interference of the excited cladding modes with the propagating light inside the core. This interference yields directed spectrum within the LPFG resonance bands envelop.

6. Conclusions

In this study, a bi-functional coated tapered LPFG sensor for gas and temperature sensitivity was fabricated and investigated. The effectiveness of the proposed methods has an impact on the sensor performance in terms of gas and temperature sensitivities which were enhanced significantly compared with other reported studies. The metal layer increases the heat conductivity, mechanical strength and temperature range of the fiber. Besides, the taper region and its structure increase the evanescent field intensity and conversion efficiency on the fiber core. However, the temperature sensitivity was calculated to be 4.9 pm/℃ under fixed NH3 concentration. The gas sensitivity of 4% NH3 at room temperature was estimated to be 0.41 nm%-1. The sensor shows a good recoverably, repeatability and stability for different applied gases concentrations and temperatures. It is important to mention that more consideration should be taken into account for the dimensions of the taper structure and the coated material thickness during the sensor fabrication process.

  References

[1] Zhan, Y., Xue, S., Yang, Q., Xiang, S., He, H., Zhu, R. (2008). A novel fiber Bragg grating high temperature sensor. Optik, 119(11): 535-539. http://dx.doi.org/10.1016/j.ijleo.2007.02.010

[2] Urban, F., Kadlec, J., Vlach, R., Kuchta, R. (2010). Design of a pressure sensor based on optical fiber Bragg grating lateral deformation. Sensors, 112(10): 12-25. http://dx.doi.org/10.3390/s101211212 

[3] Stefani, A., Yuan, W., Andresen, S., Bang, O. (2010). Polymer optical fiber Bragg grating sensors: measuring acceleration. In Proceedings of the 35th Australian Conference on Optical Fibre Technology (ACOFT) IEEE, pp. 1-4. http://dx.doi.org/10.1109/ACOFT.2010.5929947 

[4] Reck, K., Thomsen, E.V., Hansen, O. (2011). MEMS Bragg grating force sensor. Optical Express, 19(20): 19190-19198. http://dx.doi.org/10.1364/OE.19.019190 

[5] Comanici, M., Chen, L., Kung, P., Wang, L. (2011). Measurement of dynamic strain using a fiber Bragg grating-based laser sensor system. In: Proceedings of the 2011 ICO International Conference on Information Photonics (IP), pp. 1-2. http://dx.doi.org/10.1109/ICO-IP.2011.5953788 

[6] Jiang, Q., Hu, D. (2011). Micro displacement sensor based on tilted fiber Bragg grating transversal load effect. Sensor Journal IEEE; 11(9): 1776-1779. http://dx.doi.org/10.1109/JSEN.2010.2103399

[7] Ambrosino, C., Capoluongo, P., Campopiano, S., Cutolo, A., Giordano, M., Davino, D., Visone, C., Cusano, A. (2010). Fiber Bragg grating and magnetic shape memory alloy: Novel high-sensitivity magnetic Sensor. Sensor Journal IEEE, 7(2): 228-229. http://dx.doi.org/10.1109/JSEN.2006.886905

[8] Ecke, W., Schroeder, K. (2009). Fiber Bragg grating opto chemical sensor basing on evanescent-field interaction with surface plasmon waves. In: Proceedings of the 16th International Symposium on: Smart Structures and Materials and Nondestructive Evaluation and Health Monitoring. International Society for Optics and Photonics, 2019: 72930A. http://dx.doi.org/10.1117/12.816356

[9] Alshaikhli, Z.S., Hekmat, W.A. (2022). Analysis of pressure and temperature sensitivity based on coated cascade FBG-LPFG sensor. Lecture Notes in Networks and Systems, 430: 491-501. https://doi.org/10.1007/978-981-19-0825-5_53 

[10] Mohammed, N.A., Ali, T.A., Aly, M.H., Member, O.S.A. (2014). Evaluation and performance enhancement for accurate FBG temperature sensor measurement with different apodization profiles in single and quasi-distributed DWDM systems. Optics and Lasers in Engineering, 55: 22-34. https://doi.org/10.1016/j.optlaseng.2013.10.013

[11] Ma, J., Wu, S., Cheng, H., Yang, X., Wang, S., Lu, P. (2021). Sensitivity-enhanced temperature sensor based on encapsulated S-taper fiber Modal interferometer. Optics & Laser Technology, 139: 106933. https://doi.org/10.1016/j.optlastec.2021.106933

[12] Wu, Y., Liu, B., Nan, T., Wu, J., Mao, Y., Ren, J., Zhang, Y. (2021). Fiber optic hybrid structure based on an air bubble and thin taper for measurement of refractive index, temperature, and transverse load. Optik, 241: 166962. https://doi.org/10.1016/j.ijleo.2021.166962 

[13] Jin, X., Lu, C., Lin, J., Chen, X., Li, X., Xiang, Z.,Yuan, L. (2021). A strain sensor with low temperature crosstalk based on re-modulation of D-shaped LPFG. Measurement, 177: 109300. https://doi.org/10.1016/j.measurement.2021.109300

[14] Jin, X., Xiang, Z., Lu, C., et al. (2021). Highly sensitive vector bending sensor based on an embedded multimode D-shaped LPFG. Optics Express, 29(15): 22813-22822. https://doi.org/10.1364/OE.427251 

[15] Zhao, X., Zhao, N., Shi, Y., Xin, H., Li, B. (2020). Optical fiber tweezers: A versatile tool for optical trapping and manipulation. Micromachines, 11(2): 114. https://doi.org/10.3390/mi11020114 

[16] Sun, D., Ran, Y., Wang, G. (2017). Label-free detection of cancer biomarkers using an in-line taper fiber-optic interferometer and a fiber Bragg grating. Sensors, 17(11): 2559. https://doi.org/10.3390/s17112559 

[17] Wu, C. (2018). Magnetic field sensor based on nickel-coated S-shaped long period fiber grating. Optical and Quantum Electronics, 50(10): 1-11. https://doi.org/101007/s11082-018-1626-6

[18] Qi, W., Chao, Du., Jiaming, Z., Riqing, L., Yong, Z. (2016). Sensitivity-enhanced temperature sensor based on PDMS-coated long period fiber grating. Optics Communications, 377(15): 89-93. https://doi.org/10.1016/j.optcom.2016.05.039 

[19] Coelho, L., Viegas, D., Santos, J.L. (2016). Optical sensor based on hybrid FBG/titanium dioxide coated LPFG for monitoring organic solvents in edible oils. Talanta, 148(1): 170-176. https://doi.org/10.1016/j.talanta.2015.10.067

[20] Taha, B.A., Ali, N., Sapiee, N.M., Fadhel, M.M., Mat Yeh, R.M., Bachok, N.N., Al Mashhadany, Y., Arsad, N. (2021). Comprehensive review tapered optical fiber configurations for sensing application: Trend and challenges. Biosensors, 11(8): 253. https://doi.org/10.3390/bios11080253

[21] Han, M., Guo, F., Lu, Y. (2010). Optical fiber refractometer based on cladding-mode Bragg grating. Optics Letters, 35(3): 399-401. https://doi.org/10.1364/OL.35.000399

[22] Zakrzewski, A., Pięta, A., Patela, S. (2016). Simple method for manufacturing and optical characterization of tapered optical fibres. Opto−Electronics Review, 24(4): 216-222. https://doi.org/10.1515/oere-2016-0024

[23] Korposh, S., James, S.W., Lee, S.W., Tatam, R.P. (2019). Tapered optical fibre sensors: Current trends and future perspectives. Sensors, 19(10): 2294. https://doi.org/10.3390/s19102294

[24] Li, C., Yang, W., Wang, M., Yu, X., Fan, J., Xiong, Y., Yang, Y., Li, L. (2020). A review of coating materials used to improve the performance of optical fiber sensors. Sensors, 20(15): 4215. https://doi.org/10.3390/s20154215

[25] Guzowski, B., Łakomski, M., (2021). Temperature sensor based on periodically tapered optical fibers. Sensors, 21(24): 8358. https://doi.org/10.3390/s21248358 

[26] Linesh, J., Libish, T.M., Bobby, M.C., Radhakrishnan, P., Nampoori, V.P.N. (2011). Periodically tapered LPFG for ethanol concentration detection in ethanol-gasoline blend. Sensors & Transducers Journal, 125(2): 205-212. 

[27] Bock, W.J., Chen, J., Mikulic, P., Eftimov, T. (2007). A novel fiber-optic tapered long-period grating sensor for pressure monitoring. In IEEE Transactions on Instrumentation and Measurement, 56(4): 1176-1180. https://doi.org/10.1109/TIM.2007.899904